77 posts
PE102-075 flow meter added to Center Bay Leak collector
Added a Seametrics PE102-075 flow meter inline with the Leak tipping bucket in the Center Bay.
Valid flow rates for this model are 0.2 - 20 GPM.
Data for Leak tipping bucket between 11/15/2016 11:40 - 13:06 are not valid
20161110100054 XWiki.eahunt
Vacuum started at 0930 on all samplers for all three hill slopes. Vacuum stopped at 0630 next morning. Collected all sample available as Set 1 for geochemistry.
LEO bigPERTH experiment started
The LEO bigPERTH experiment was started on Nov. 6, 2016 at 23:00 MST.
Irrigation actuators on LEO East and West did not respond, so circuits were opened and closed in manual mode with a slight start-time delay (1-5 minutes).
Irrigation valve actuator swap 2016117093157
On Nov. 6, 2016 from about 10:45 PM - 11:30 PM Aaron and Michael performed work on the electric actuators controlling/monitoring LEO irrigation.
Several of the actuators were not responding (No diagnostics have been performed), so we swapped around actuators such that each bay now should have a functional actuator on both the MULTI1 and MULTI2 circuit.
Due to this change several irrigation circuits will display as "in transition" because they have been physically removed / electrically disconnected.
SDI-12 sensors removed
Today several more SDI-12 sensors were removed from the bus due to large power draw in:
East bay zones P & D
Center bay zone K
LEO-C_4_0_2_HMP60 malfunction
The HMP60 in the center bay, mast 4,0 position 2 appears to be incorrectly reporting RH. Output signal has been verified and matches recorded value.
LEO SDI-12 system powered on for cooling experiment
The SDI-12 systems in all three bays were powered on to support an air flow / cooling experiment. VWC is required for the surface heat flux computation.
LEO Sampling Interval Change
July 24, 2016
The sampling interval for most LEO sensors, excluding 5TM, MPS-2, and CS451, has been set to 1 minute for a series of above slope air flow experiments. Since there is no anticipated irrigation I have decided to leave the SDI-12 sensors (listed above) at the standard 15 minute logging interval.
Mini LEO calibrations 20160609
Antonio has performed calibrations for the drainage tipping bucket and EC meter. As of today (06/09/2016) the scales for these two devices have been updated on the data logger.
TB - 0.007007 liters per tip
EC = 711.167 * voltage - 124.152