Overview

Honeywell Model3130 Load Cell (Manufacturers webpage)

Calculation Notes

Sensor Code:Model3130
Sensor:Model3130 from Honeywell International Inc.
Measurement:Force (converted to mass in kg)
Output 1:mV/V (unit-less)Range 1:-0.003 to 0.003
Output 2:Mass (kg) = Output1*\frac{{MaxLoad} * {453.59236}}{k}Range 2:0 to 2*108

Calibration coefficient, k, is the output at the rated MaxLoad (kip). For mass calculations: m=F/a, where a is 9.80665 (gravity) and 1 kip = 4448.2216 newtons (N)

Wiring Diagram

Model3130_Wiring_Schematic.png

Tags:
  
This wiki is licensed under a Creative Commons 2.0 license