Michael Sibayan 1.1 1 CSAT3 control box has been modified with an RS232 to RS422 transceiver to allow for a longer transmission line at 19.2 kbps
2 Using USConverters.com device, however, any comparable converter should work.
3
Wei-Ren Ng 3.3 4 The CSAT3 has to be configured in the Unprompted mode with the synchronization code enabled due to RS422 not utilizing RTS/CTS or DTR/DSR.
Wei-Ren Ng 2.2 5
6 CSAT3 is set to the following configuration with the corresponding commands:
Michael Sibayan 1.1 7 rs 1 - AA55 synchronization code
Wei-Ren Ng 2.2 8 br 1 - Baud rate 19.2 kbps
Michael Sibayan 1.1 9 ri 1 - RTS independent
Wei-Ren Ng 2.2 10 & - Unprompted mode
11 Ae - 60 Hz Execution Parameter/Acquisition rate.
Michael Sibayan 1.1 12
Wei-Ren Ng 2.2 13 Settings are saved to EEPROM by the 'sr2718' command. 
Wei-Ren Ng 3.1 14 *Do note that a hardware jumper needs to be set in the electronics box. Refer to manual for details.
Michael Sibayan 1.1 15
Wei-Ren Ng 3.3 16 The total output per measurement is 6 words including the 'AA55' sync code.
17
Wei-Ren Ng 3.2 18 EOL by Campbell Sci is December 2026.
Wei-Ren Ng 3.3 19
20
21 The default buffer size of the cRIO controllers is 4096 Bytes. Currently, the cRIO is programmed to read 720 Bytes or 60 measurements on the COM port. However, only the first measurement or first 12 Bytes is processed to update the shared variable every second.