1 [[image:cRIO-TRF-P01.jpg]]
2
3 __Module 1 (NI-9401)__
4 * 8 channel digital I/O set up as a pulse counter.
5 * Used to monitor pulses from anemometers
6 * Terminal block in the upper right provides power (through a resistor) to the anemometers.
7
8
9 __Module 3 (NI-9205)__
10 * 16/32 channel +/-10V analog input
11 * First 16 channels set up in +/-1V differential mode for the pyranometers and quantum sensors
12 * Last 16 channels set up in +/-5V referenced single ended mode for the HMP45 temp/RH sensors
13 * Common is tied to the 12VDC power supply negative terminal (for RSE channels)
14
15 __Module 5 (NI-9213)__
16 * 16 channel +/- 75mV differential analog input
17 * First four channels measure light sensor signals
18 * Channel 5 measures voltage of the PT100 thermistor
19 * Common is tied to the shields and the 12VDC power supply negative terminal (for PT100)
20
21
22 __0.5mA current source__
23 * Provides approximately constant 0.5mA current for the PT-100 thermistor
24 * 0.5mA chosen to put the signal voltage within range of the NI-9213 AI module (approximately 50mV)
25
26
27 __Anemometer terminal block__
28 * Provides 5VDC, through a resistor, to the anemometers
29 * 5V chosen to match the TTL level of the NI-9401 DIO module